Title of article :
Accuracy of Cerenkov measurements using a liquid scintillation spectrometer
Author/Authors :
Makoto Takiue، نويسنده , , Takashi Natake، نويسنده , , Haruo Fujii، نويسنده , , Tamaru Aburai، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
4
From page :
123
To page :
126
Abstract :
Cerenkov counting efficiency varies with colour quenching and sample turbidity. The activity in a plastic vial can be determined accurately with a colour quenching correction technique, regardless of the presence or absence of turbidity in a sample. On the other hand, the error of the measured activity in a glass vial becomes large with increasing sample turbidity due to the dissimilarity of the quench correction curves for non-turbid and turbid samples.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1996
Journal title :
Applied Radiation and Isotopes
Record number :
539443
Link To Document :
بازگشت