• Title of article

    Selective saturation method for EPR dosimetry with tooth enamel

  • Author/Authors

    E. A. Ignatiev، نويسنده , , A. A. Romanyukha، نويسنده , , A. A. Koshta، نويسنده , , A. Wieser، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    5
  • From page
    333
  • To page
    337
  • Abstract
    The method of selective saturation proposed is based on the difference in the microwave (mw) power dependence of the background and radiation induced EPR components of the tooth enamel spectrum. The substraction of the EPR spectrum recorded at low mw power from that recorded at higher mw power provides a considerable reduction of the background component in the spectrum. The resolution of the EPR spectrum could be improved 10-fold, however simultaneously the signal-to-noise ratio was found to be reduced twice. A detailed comparative study of reference samples with known absorbed doses was performed to demonstrate the advantage of the method. The application of the selective saturation method for EPR dosimetry with tooth enamel reduced the lower limit of EPR dosimetry to about 100 mGy.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1996
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    539478