Title of article :
Dependence of photon interaction depth on linear attenuation coefficient in high pure germanium detectors
Author/Authors :
Haluk Yucel، نويسنده , , M. Atif Cetimagener، نويسنده , , Halimagel Demimagerel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Abstract :
The present work is concerned with the construction of a semiempirical function de(μ, Eγ) of the effective photon interaction depth depending upon both the linear attenuation coefficient of Ge and γ-ray energy. The interaction depth results obtained from the measurements in two high purity germanium crystals, which are nearly equal in volume, have been fitted by using a function. The semiempirical function agrees with the measured interaction depth values on the average to within 3–4%. The knowledge of effective interaction depth of γ-rays in germanium detector is a useful parameter in order to calculate the absolute efficiency for any source-to-detector distance on the particular detector.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes