Abstract :
The primary goal of radiation protection in decommissioning and decontamination of the old nuclear facilities of the CIEMAT is to monitor and minimize exposure of personnel. Monitoring programs include determination of actinides and 90Sr in biological samples. A technique for the sequential measurement of low levels of 239Pu, 241Am and 90Sr in urine samples has been developed. The method involves coprecipitation of these radionuclides as phosphates from bulk urine sample. Separation of Plutonium is carried out using a conventional anion exchange technique. Americium and strontium isolations are achieved sequentially by chromatographic extraction (Tru.Spec and Sr.Spec columns) from the load and rinse solutions coming from the anion exchange column. Plutonium and Americium measurements are performed by alpha spectrometry. The mean recovery obtained is 80% and the detection limit for 24 h urine sample (1.41) is 0.6 mBq L−1. 90Sr determination is made by liquid scintillation counting. The detection limit in this case is 1.1 E-01 Bq/L.