Title of article :
Alanine-ESR dosimetry for radiotherapy IAEA experience
Author/Authors :
Kishor Mehta، نويسنده , , Reinhard Girzikowsky، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
3
From page :
1189
To page :
1191
Abstract :
At present, the most commonly used transfer dosimeters for radiotherapy applications are TL dosimeters. They are being used for intercomparison between SSDLs (about 70) and the IAEA dosimetry laboratory. However, there are some undesirable characteristics of this dosimetry system. We have a study in progress at the IAEA to evaluate the alanine-ESR system as an alternative to TLDs. There are several desirable qualities which make alanine an attractive dosimeter. Preliminary data suggest that the alanine-ESR dosimetry system has the potential to replace TLDs for intercomparison amongst SSDLs in the therapy-level dose region.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1996
Journal title :
Applied Radiation and Isotopes
Record number :
539620
Link To Document :
بازگشت