The extension of the range of NIM alanine/ESR dosimetric system to therapy level
Author/Authors :
Gao Juncheng، نويسنده , , Wang Zaiyong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
4
From page :
1193
To page :
1196
Abstract :
The NIM alanine/ESR dosimetric system, which was designed for industrial radiation processing, has been improved to be suited for applications in therapy dose range. Two different procedures of dose intercomparisons between IAEA and NIM were carried out with the improved system in the range of 2.5 to 100 Gy. In the first procedure, a set of NIM alanine dosimeters were irradiated at IAEA dosimetric laboratory and part of dosimeters marked “for evaluated” were evaluated using the rest of those with “known dose” given by IAEA. Most of evaluated doses agreed with IAEA doses within 1%. In the second procedure, all above dosimeters were evaluated on the base of NIM dosimetry. The results indicated that the doses determined by NIM agreed with those given by IAEA within 3% on the average.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1996
Journal title :
Applied Radiation and Isotopes
Record number :