Title of article
Proton dosimetry in bone using electron spin resonance
Author/Authors
John F. Copeland، نويسنده , , Kenneth P. Gall، نويسنده , , Sang-Yeob Lee، نويسنده , , George E. Chabot، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
6
From page
1533
To page
1538
Abstract
We have studied the ESR response of proton-irradiated (in vitro) bone. The ESR response as a function of proton (E = 105 MeV) dose to bone was linear from 0 to 50 Gy and similar to the photon (E = 6 MV) dose response. The ESR depth response (Bragg) curve was depressed as compared to a depth-response curve determined with a parallel plate ionization chamber (PPIC). There was a short-term ESR signal fade in the Bragg peak region, likely attributable to the organic component in bone. We are continuing to investigate these latter two effects.