Title of article :
Automation of a pneumatic system by controlling a microcomputer equipped with a custom add on board for neutron activation analysis
Author/Authors :
M. A. Cetiner، نويسنده , , H. Demirel، نويسنده , , A. Demirbaimage، نويسنده , , H. Yucel، نويسنده , , I. Celenk، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
6
From page :
397
To page :
402
Abstract :
An automated sample transfer system associated with a 3 × 592 GBq Am---Be neutron source has been constructed at the Ankara Nuclear Research and Training Center (ANRTC) for Instrumental Neutron Activation Analysis (INAA) and Cyclic Instrumental Neutron Activation Analysis (CINAA). The pneumatic transfer system is controlled by an IBM PS/1 personal computer which is equipped with a custom add-on board for relevant hardware and control software written in Turbo Pascal. The system computer interactively works with a Multichannel Analyzer (MCA) for proper data acquisition. The sample transfer system also runs manually. The design and performance of the hardware and software are discussed.