Title of article :
The electron-dose distribution surrounding an 192Ir wire brachytherapy source investigated using EGS4 simulations and GafChromicTM film
Author/Authors :
Y. C. Cheung، نويسنده , , Peter K. N. Yu، نويسنده , , E. C. M. Young، نويسنده , , C. L. Chan، نويسنده , , M. F. Ng، نويسنده , , Francis N. F. Tang، نويسنده , , Tony P. Y. Wong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
6
From page :
985
To page :
990
Abstract :
The steep dose gradient around 192Ir brachytherapy wire implants is predicted by the EGS4 (PRESTA version) Monte Carlo simulation. When considering radiation absorbion regions close to the wire source, the accurate dose distribution cannot be calculated by the GE Target II Sun Sparc treatment-planning system. Experiments using GafChromicTM film have been performed to prove the validity of the EGS4 user code when calculating the dose close to the wire source in a low energy range.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1997
Journal title :
Applied Radiation and Isotopes
Record number :
539859
Link To Document :
بازگشت