• Title of article

    Measurement standards for strontium-89 for use in bone palliation

  • Author/Authors

    B. M. Coursey، نويسنده , , F. J. Schima، نويسنده , , D. B. Golas، نويسنده , , O. T. Palabrica، نويسنده , , A. Suzuki، نويسنده , , M. A. Dell، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    10
  • From page
    335
  • To page
    344
  • Abstract
    Strontium-89 standards have been prepared for use in calibrating instruments in the measurement chain from production in reactors to administration in the clinic or radiopharmacy. Alternate reactor production schemes were evaluated to yield high purity 89Sr with minimum 85Sr impurity. Following purification to remove radionuclidic impurities, samples of 89Sr were standardized by high-efficiency liquid-scintillation counting with a relative expanded uncertainty (intended to approximate two standard deviations) of 0.48%. A Standard Reference Material, SRM 4426A, was prepared and distributed to radiopharmaceutical manufactures and other customers. The standard sources of 89Sr were used in different geometries to calibrate high purity Ge semiconductor detectors, re-entrant ionization chambers and commercial radionuclide calibrators. The latter included Capintec dose calibrators and the Capintec βeta C™ NaI(Tl) scintillation counter.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1998
  • Journal title
    Applied Radiation and Isotopes
  • Record number