Title of article :
Improvements to the in vivo measurement of cadmium in the kidney by neutron activation analysis
Author/Authors :
F. E. McNeill، نويسنده , , D. R. Chettle، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
2
From page :
699
To page :
700
Abstract :
The McMaster University in vivo cadmium measurement system has been upgraded. Two large surface area hyperpure germanium planar detectors (51 mm diameter × 21 mm depth) are now employed for the detection of the 559 KeV γ-signal from cadmium. Phantom studies indicate that the new detectors have significantly improved the system. The detection limit in kidney phantoms has been reduced from 8.9 to 3.8 mg for the same dose and measurement time.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1998
Journal title :
Applied Radiation and Isotopes
Record number :
540107
Link To Document :
بازگشت