• Title of article

    Improvements to the in vivo measurement of cadmium in the kidney by neutron activation analysis

  • Author/Authors

    F. E. McNeill، نويسنده , , D. R. Chettle، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    2
  • From page
    699
  • To page
    700
  • Abstract
    The McMaster University in vivo cadmium measurement system has been upgraded. Two large surface area hyperpure germanium planar detectors (51 mm diameter × 21 mm depth) are now employed for the detection of the 559 KeV γ-signal from cadmium. Phantom studies indicate that the new detectors have significantly improved the system. The detection limit in kidney phantoms has been reduced from 8.9 to 3.8 mg for the same dose and measurement time.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1998
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    540107