Title of article :
An analyzer for pulse-interval times to study high-order effects in the processing of nuclear detector signals
Author/Authors :
B. Denecke، نويسنده , , S. de Jonge، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
7
From page :
1099
To page :
1105
Abstract :
An electronic device to measure interval time density distributions of subsequent pulses in nuclear detectors and their electronics is described. The device has a pair-pulse resolution of 10 ns and 25 ns for 3 subsequent input signals. The conversion range is 4096 channels and the lowest channel width is 10 ns. Counter dead times, single and in series were studied and compared with the statistical model. True count rates were obtained from an exponential fit through the interval–time distribution.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1998
Journal title :
Applied Radiation and Isotopes
Record number :
540172
Link To Document :
بازگشت