Title of article :
Standardization of 110mAg and 75Se by the beta-efficiency extrapolation method
Author/Authors :
E.L. Grigorescu، نويسنده , , Maria Sahagia، نويسنده , , Anamaria Razdolescu، نويسنده , , A. Luca، نويسنده , , R.M. Radwan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
The complex beta–gamma emitter 110mAg and the electron-capture radionuclide 75Se were standardized, using the “beta-efficiency extrapolation method”. The coincidence equipment was a NE modular type. The detectors were a NaI (Tl) counter and a 4π proportional counter. For 110mAg, one tried to fulfil the “linearity conditions”. For 75Se one tried to minimize errors arising from an isomeric state and strong internal conversion; an “equivalent scheme” was used. Combined uncertainties of ≈1% were obtained.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes