• Title of article

    Calculation of the average path of gamma-rays in the crystal of semiconductor gamma-ray spectrometers using the total efficiency calibration curve

  • Author/Authors

    Matja? Korun، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    4
  • From page
    1231
  • To page
    1234
  • Abstract
    A method for calculating the average path traversed in the detector crystal by the detected gamma-rays using the total efficiency calibration curves for point sources placed close to the detector was developed. The influence of the crystal volume, shape and dead layer on the average path is discussed. A method for extrapolation of the total efficiency calibration curves to energies up to 5 MeV is presented.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1998
  • Journal title
    Applied Radiation and Isotopes
  • Record number