Title of article
Intercomparisons of 67Ga and 123I assays in UK hospitals, 1996
Author/Authors
M.J. Woods، نويسنده , , J.D Keightley، نويسنده , , M. Ciocanel، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
4
From page
1449
To page
1452
Abstract
As part of its ongoing support to the National Measurement System in the UK, the National Physical Laboratory (NPL) conducted two intercomparison exercises in 1996. The participants, hospital physicists, were required to assay samples of 67Ga and 123I using radionuclide calibrators. The analysis of the results presented here, whilst encouraging in terms of overall level of quality compared with previous exercises, identified several areas of concern. The results are discussed, conclusions drawn and recommendations made.
Journal title
Applied Radiation and Isotopes
Serial Year
1998
Journal title
Applied Radiation and Isotopes
Record number
540235