• Title of article

    Dosimetry for optimisation of patient protection in computed tomography

  • Author/Authors

    K. A. Jessen، نويسنده , , P. C. Shrimpton، نويسنده , , J. Geleijns، نويسنده , , W. Panzer، نويسنده , , G. Tosi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    8
  • From page
    165
  • To page
    172
  • Abstract
    The complex conditions of irradiation in computed tomography (CT), involving highly-collimated X-ray beams, necessitate the use of specially-defined dose descriptors such as the computed tomography dose index (CTDI). When used in a weighted form (CTDIW), this concept can describe the absorbed dose from a single slice in standard head and body phantoms. The model can easily be extended to characterise patient exposure for a complete examination by means of the reference dose quantity dose-length product (DLP). Effective dose can also be estimated from DLP, when required.
  • Keywords
    radiation dose , dosimetry , computer assisted tomography
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1999
  • Journal title
    Applied Radiation and Isotopes
  • Record number