Title of article :
Reproducibility study of TLD-100 micro-cubes at radiotherapy dose level
Author/Authors :
Luiz Antonio R. da Rosa، نويسنده , , Dieter F. Regulla، نويسنده , , Ute A. Fill، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
5
From page :
573
To page :
577
Abstract :
The precision of the thermoluminescent response of Harshaw micro-cube dosimeters (TLD-100), evaluated in both Harshaw thermoluminescent readers 5500 and 3500, for 1 Gy dose value, was investigated. The mean reproducibility for micro-cubes, pre-readout annealed at 100°C for 15 min, evaluated with the manual planchet reader 3500, is 0.61% (1 standard deviation). When micro-cubes are evaluated with the automated hot-gas reader 5500, reproducibility values are undoubtedly worse, mean reproducibility for numerically stabilised dosimeters being equal to 3.27% (1 standard deviation). These results indicate that the reader model 5500, or, at least, the instrument used for the present measurements, is not adequate for micro-cube evaluation, if precise and accurate dosimetry is required. The difference in precision is apparently due to geometry inconsistencies in the orientation of the imperfect micro-cube faces during readout, requiring careful and manual reproducible arrangement of the selected micro-cube faces in contact with the manual reader planchet.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1999
Journal title :
Applied Radiation and Isotopes
Record number :
540346
Link To Document :
بازگشت