Title of article :
Influence of radioactive impurities on SIR measurements
Author/Authors :
C. Michotte، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
319
To page :
323
Abstract :
The correction usually applied to an SIR measurement in the presence of a gamma-ray emitting impurity is presented and explained. The method of calibration of a Ge(Li) spectrometer is briefly described. It is used to look for impurities when SIR measurements show inconsistencies. An example is presented where a 103Ru impurity was discovered in a 99Mo solution. The correction applied compensates for the decrease of around 0.2% per day observed in the SIR measurements of this 99Mo ampoule.
Journal title :
Applied Radiation and Isotopes
Serial Year :
2000
Journal title :
Applied Radiation and Isotopes
Record number :
540551
Link To Document :
بازگشت