Title of article :
The standardization of a 204Tl solution
Author/Authors :
Maria Sahagia، نويسنده , , Anamaria Razdolescu، نويسنده , , E. L. Grigorescu، نويسنده , , A. Luca، نويسنده , , C. Ivan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
Two methods were used for the standardization of the 204Tl solution received in the framework of the 1997 BIPM comparison. The first method, 4π liquid scintillation (LS), consisted of the extrapolation of the 4π counting rate to the zero discrimination level. A home-made, single channel instrument and a commercial Packard INSTA-GEL scintillator were used. The radioactivity concentration on the reference date 1 April 1997 was 63.1±1.3 kBq g−1. The second method was the efficiency tracer technique, using a 60Co standard solution; the radioactivity concentration was calculated as the mean of several extrapolated values of the curve NβTl-204=N0Tl-204f(1−var epsilonβCo-60).
The preliminary reported value was 62.8±0.8 kBq g−1; the revised value is 61.8±1.2 kBq g−1, in good agreement with other 4π (p.c.) tracer results in the comparison. Some remarks are made regarding our measurements and the comparison results.
Keywords :
204Tl solution , liquid scintillation , Efficiency tracer , 60Co standard solution , BIPM comparison
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes