• Title of article

    MAC3: an electronic module for the processing of pulses delivered by a three photomultiplier liquid scintillation counting system

  • Author/Authors

    J. Bouchard، نويسنده , , P. Cassette، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    4
  • From page
    669
  • To page
    672
  • Abstract
    In the field of radioactivity measurement, there is a constant need for high quality electronic modules such as ADCs, amplifiers, high voltage generators, dead-time modules, etc. and sometime there is a need for a specialized module not available on the market. The purpose of the module presented here, called MAC3 (module dʹacquisition de coïncidences triples), is to process the pulses delivered by the three detectors (photomultipliers) used in our liquid scintillation counting system. The dead-time generated by this module is of the extendible type and the dead-time corrections are made according to the live-time method. This module, which has been developed and tested at LPRI, can replace the complex interconnection of several independent and costly modules. The philosophy governing the choice and the implementation of this type of dead-time as well as the system used for the dead-time corrections is presented. The electronic scheme and the performances are also presented. This module is available in the NIM standard.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2000
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    540606