Title of article :
A method for the selection of the optimum counting conditions in a ZnS(Ag) scintillation detector
Author/Authors :
Ll. Pujol، نويسنده , , J. A. Suarez-Navarro، نويسنده , , M. Montero، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
The well-known criteria for the selection of the optimum counting conditions in a ZnS(Ag) scintillation detector seem to restrict its applicability. These do not consider simultaneously operating voltage and electronic threshold variation to reach the best counting conditions. Therefore, a more general method for the determination of the optimum counting conditions in ZnS(Ag) scintillation detectors is proposed. In this method, a relationship between voltage and electronic threshold is derived for counting efficiency. In order to test the method, quality control procedures have been carried out as well as the determination of gross alpha activity in environmental samples.
Keywords :
environmental radioactivity , ZnS(Ag) scintillation detector , Gross alpha , Optimum counting conditions
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes