Title of article :
Electron arc therapy treatment planning verification with alanine/EPR dosimetry
Author/Authors :
C. De Angelis، نويسنده , , A. Mattacchioni، نويسنده , , S. Onori، نويسنده , , D. Aragno، نويسنده , , U. de Paula، نويسنده , , V. Panichelli، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
1203
To page :
1207
Abstract :
Alanine/EPR dosimetry was applied to measure dose distributions during electron arc therapy treatments generated by electron beams moving isocentrically over a stated arc. Alanine-polyethylene pellets, prepared at ISS according to the NIST recipe, inserted in a homogeneous cylindrical and anthropomorphic phantoms, were used. Preliminary, alanine response to static electron beams in the (6–20) MeV nominal energy range was studied. Then alanine dosimetry was applied to determine the dose versus the gantry angle at a reference point in a cylindrical homogeneous phantom. Finally, arc therapy treatment planning verification was performed in anthropomorphic phantom.
Journal title :
Applied Radiation and Isotopes
Serial Year :
2000
Journal title :
Applied Radiation and Isotopes
Record number :
540680
Link To Document :
بازگشت