Title of article :
Low level gamma spectrometry by beta–gamma coincidence
Author/Authors :
A. Luca، نويسنده , , P. de Felice، نويسنده , , G. Tanase، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A beta–gamma coincidence system consisting of a 2π plastic beta detector and a Ge(Li) detector, is described. Such a system allows for a drastic reduction of background (not, vert, similar70 times) as well as of Minimum Detectable Activity (not, vert, similar7 times). An integral background (50–1500 keV) of less than 0.1 cps is obtained. The difficulty of the method, otherwise simple and inexpensive, is the precise determination of beta detector efficiency.
Keywords :
Gamma Spectrometry , Beta counters
Journal title :
Applied Radiation and Isotopes
Journal title :