Title of article :
Computed tomography with monochromatic bremsstrahlung radiation
Author/Authors :
R. T. Lopes، نويسنده , , E. B. Costa، نويسنده , , E. F. O. De Jesus، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
7
From page :
665
To page :
671
Abstract :
The use of filters with appropriate thickness of different elements produces a method to quasi-monochromatize X-ray beams. In the present study the filters were produced in a boric acid matrix, using the elements Mo, Sb, Cs, Ce and Gd. The thickness of the filters varied from 50 to 300 μm for a high voltage tube operating from 30 to 70 kVp. The parameters analyzed were the beam resolution, measured by its FWHM as a function of filter thickness, and the reduction of beam intensity. The results show a comparison between conventional tomographies obtained with different filters and application in the technique of differential tomography, with double energy for identification of elements such as Zr, Nd, Ba and Sm.
Keywords :
computed tomography , Bremsstrahlung radiation , Monoenergetic radiation
Journal title :
Applied Radiation and Isotopes
Serial Year :
2000
Journal title :
Applied Radiation and Isotopes
Record number :
540814
Link To Document :
بازگشت