Calibration and characterization of a digital X-ray fluorescence bone lead system
Author/Authors
David E. B. Fleming، نويسنده , , Thomas A. Forbes، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
6
From page
527
To page
532
Abstract
Five different combinations of digital shaping parameters were tested for a newly assembled, 109Cd source, K X-ray fluorescence bone lead system. System calibration results are presented, along with analyses of measurement uncertainty and reproducibility obtained from repeat measurements of a bone phantom and a human tibia. Digital shaping parameters of 2.4 μs for a rise time/fall time and 1.2 μs for a flat top width were identified as superior. The digital system provided significant improvements in overall measurement precision, with gains of at least 25–35% over conventional system results.
Keywords
bone lead , X-RAY FLUORESCENCE , Digital spectroscopy