• Title of article

    Calibration and characterization of a digital X-ray fluorescence bone lead system

  • Author/Authors

    David E. B. Fleming، نويسنده , , Thomas A. Forbes، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    527
  • To page
    532
  • Abstract
    Five different combinations of digital shaping parameters were tested for a newly assembled, 109Cd source, K X-ray fluorescence bone lead system. System calibration results are presented, along with analyses of measurement uncertainty and reproducibility obtained from repeat measurements of a bone phantom and a human tibia. Digital shaping parameters of 2.4 μs for a rise time/fall time and 1.2 μs for a flat top width were identified as superior. The digital system provided significant improvements in overall measurement precision, with gains of at least 25–35% over conventional system results.
  • Keywords
    bone lead , X-RAY FLUORESCENCE , Digital spectroscopy
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2001
  • Journal title
    Applied Radiation and Isotopes
  • Record number