Title of article
Transfer of the efficiency calibration of Germanium gamma-ray detectors using the GESPECOR software
Author/Authors
O. Sima، نويسنده , , Eugene D. Arnold، Jr نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
5
From page
71
To page
75
Abstract
The GESPECOR software was extended to incorporate procedures for the computation of the efficiency transfer factor for cases of practical interest: (a) sources with identical geometry, but different matrices; (b) sources with similar (but not identical) geometry; and (c) transfer from a point source to a volume source. Fast and accurate algorithms, based on correlated sampling, were implemented for solving the first two cases. A procedure to take into account the imperfect charge collection in the detector was implemented.
Keywords
Efficiency transfer , Matrix effects , Geometry effects , GESPECOR , Gamma-spectrometry