Title of article :
Fitting methods for constructing energy-dependent efficiency curves and their application to ionization chamber measurements
Author/Authors :
A. Svec، نويسنده , , H. Schrader، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
An ionization chamber without and with an iron liner (absorber) was calibrated by a set of radionuclide activity standards of the Physikalisch-Technische Bundesanstalt (PTB). The ionization chamber is used as a secondary standard measuring system for activity at the Slovak Institute of Metrology (SMU). Energy-dependent photon-efficiency curves were established for the ionization chamber in defined measurement geometry without and with the liner, and radionuclide efficiencies were calculated. Programmed calculation with an analytical efficiency function and a nonlinear regression algorithm of Microsoft (MS) Excel for fitting was used. Efficiencies from bremsstrahlung of pure beta-particle emitters were calibrated achieving a 10% accuracy level. Such efficiency components are added to obtain the total radionuclide efficiency of photon emitters after beta decay. The method yields differences of experimental and calculated radionuclide efficiencies for most of the photon-emitting radionuclides in the order of a few percent.
Keywords :
Analytical function , ?-connection , calibration , Radionuclide activity , ionization chamber , Optimized parameters , Efficiency curve
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes