Title of article :
A new absolute method for the standardization of radionuclides emitting low-energy radiation
Author/Authors :
E. Leblanc، نويسنده , , P. de Marcillac، نويسنده , , N. Coron، نويسنده , , J. Leblanc، نويسنده , , M. Loidl، نويسنده , , J. F. Metge، نويسنده , , J. Bouchard، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
245
To page :
251
Abstract :
Microcalorimeters (or bolometers) operated at temperatures below 100 mK allow individual counting of photons and electrons with a very low energy detection threshold. The physics is based on the pulse temperature increase of the target (or absorber) of the detector due to the complete absorption of both electrons and photons. Since this target can be constructed with a perfect 4-π geometry, a bolometer offers potentially a new method for absolute activity measurements of radionuclides emitting low-energy radiation. In this paper we present our first results of a feasibility study of activity standardization of a 55Fe solution with a prototype 4-π bolometer.
Journal title :
Applied Radiation and Isotopes
Serial Year :
2002
Journal title :
Applied Radiation and Isotopes
Record number :
541154
Link To Document :
بازگشت