Title of article :
Determination of the dosimetric characteristics of InterSource125 Iodine brachytherapy source
Author/Authors :
Ali S. Meigooni، نويسنده , , Maung M. Yoe-Sein، نويسنده , , Awni Y. Al-Otoom، نويسنده , , Keith T. Sowards، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
11
From page :
589
To page :
599
Abstract :
The TG-43 recommended dosimetric characteristics of a new 125I brachytherapy source have been experimentally and theoretically determined. The measurements were performed in Solid Water™ using LiF TLDs. The calculations were performed using Monte Carlo simulations in Solid Water™ and water. The measured data were compared with calculated values as well as the reported data in literature for other source designs. The dose rate constant this source in water was 1.01±3% cGy h−1 U−1 and the anisotropy constant was 0.956.
Keywords :
TLD , brachytherapy , thermoluminescent dosimeter , dosimetry , TG-43 , 125I