• Title of article

    Determination of the dosimetric characteristics of InterSource125 Iodine brachytherapy source

  • Author/Authors

    Ali S. Meigooni، نويسنده , , Maung M. Yoe-Sein، نويسنده , , Awni Y. Al-Otoom، نويسنده , , Keith T. Sowards، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    11
  • From page
    589
  • To page
    599
  • Abstract
    The TG-43 recommended dosimetric characteristics of a new 125I brachytherapy source have been experimentally and theoretically determined. The measurements were performed in Solid Water™ using LiF TLDs. The calculations were performed using Monte Carlo simulations in Solid Water™ and water. The measured data were compared with calculated values as well as the reported data in literature for other source designs. The dose rate constant this source in water was 1.01±3% cGy h−1 U−1 and the anisotropy constant was 0.956.
  • Keywords
    TLD , brachytherapy , thermoluminescent dosimeter , dosimetry , TG-43 , 125I
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2002
  • Journal title
    Applied Radiation and Isotopes
  • Record number