Title of article :
Precise Monte Carlo simulation of gamma-ray response functions for an NaI(Tl) detector
Author/Authors :
Hu-Xia Shi، نويسنده , , Bo-Xian Chen، نويسنده , , Ti-Zhu Li، نويسنده , , Di-Yun Ruan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
Three widely used methods to calculate the response functions for NaI(Tl) detectors were investigated. The methods were Berger–Seltzerʹs method (Nucl. Instrum. Methods 104 (1972) 317), general Monte Carlo (MC) programs, such as EGS4 (The EGS4 code system, Stanford Linear Accelerator Center, 1985) and MCNP4B (MCNP—a general Monte Carlo N-particle transport code, Los Alamos National Laboratory Report, LA-12625-M, 1997), and special MC programs. The pulse height spectra in a 3″×3″ NaI(Tl) detector due to several gamma-ray sources have been measured to verify the calculated results of these methods. The energies of the sources ranged from 0.4118 to 7.11 MeV. The spectra generated by Berger–Seltzerʹs method and the general MC programs did not agree well with the experimental data. PETRANS 1.0, the special MC program developed in house, was fairly accurate since it also considered the scintillation efficiency and the single escape peak shift.
Keywords :
NaI(Tl) detector , Precise response function , Monte Carlo simulation , Single escape peak shift , Scintillation efficiency
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes