Absolute calibration of 60Co by using sum-peak method and an HPGe detector
Author/Authors
I. J. Kim، نويسنده , , C. S. Park، نويسنده , , H. D. Choi، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
7
From page
227
To page
233
Abstract
The sum-peak method was applied to calibrate four 60Co sources in the range of 25–350 kBq. An HPGe-based γ-ray spectroscopy system was used, and the effects of the angular correlation and pulse pile-up were corrected, and are discussed. The correction for the angular correlation was obtained from a theoretical calculation, while the pulse pile-up effect was corrected using a new extrapolation technique based on the pile-up measurements by varying the shaping time of the amplifier. The determined radioactivities were within 2% deviation from the values reported by the supplier and the associated uncertainties were less than 2%. The observed consistency was obtained under the experimental condition of keeping the total counting rate less-than-or-equals, slant15 kcps.
Keywords
Pulse pile-up , Angular correlation , Shaping time , Sum-peak method , HPGe , 60Co