Title of article :
166Ho-coated balloon for vascular brachytherapy
Author/Authors :
Young-Don Hong، نويسنده , , Sun-Ju Choi، نويسنده , , Kyung-Hwa Kim، نويسنده , , Byung-Chul Shin، نويسنده , , Woong-Woo Park، نويسنده , , Sanghyun Park، نويسنده , , Kyung Bae Park، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
8
From page :
455
To page :
462
Abstract :
A simple procedure was developed for coating the surface of a conventional percutaneous transluminal coronary angioplasty balloon angioplasty catheter with 166Ho. The absorbed dose delivered by the surface-coated catheter is twice that of a similar catheter filled with a solution containing the same activity of 166Ho. Leakage tests show that <0.6% of the coated activity is removable from the surface of the catheter. The coated catheter removes the risk of release of the 166Ho into the body as a result of rupture of the balloon, and also reduces the radiation exposure to staff.
Keywords :
? emitter , Catheter-based radiotherapy , Radioactive-coated balloon , Radiation brachytherapy , Holmium-166
Journal title :
Applied Radiation and Isotopes
Serial Year :
2003
Journal title :
Applied Radiation and Isotopes
Record number :
541453
Link To Document :
بازگشت