Title of article
Calibration of diffusion barrier charcoal detectors using a semi-empirical expression
Author/Authors
M. E. Montero Cabrera، نويسنده , , L. Colmenero Sujo، نويسنده , , M. L. Villalba، نويسنده , , J. S?enz Peinado، نويسنده , , A. Cano Jiménez، نويسنده , , A. L?pez Miranda، نويسنده , , E. F. Herrera Peraza، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
7
From page
281
To page
287
Abstract
Several calibration settings of diffusion barrier charcoal canister (DBCC) detectors for measuring radon concentration in air were studied. A set of functions and graphs were developed for relations between radon concentration in air and adsorbed activity in DBCC, when calibrated in small chambers. Both the integration time for 10% of DBCC of a batch, and the radon adsorption coefficient for the activated charcoal used in these detectors, were determined. Thus, a semi-empirical expression for detector calibration was applied.
Keywords
Diffusion barrier , calibration , Radon adsorption–desorption , Activated charcoal detectors , RADON
Journal title
Applied Radiation and Isotopes
Serial Year
2003
Journal title
Applied Radiation and Isotopes
Record number
541533
Link To Document