Title of article :
Measurement of peak and total efficiencies of low-energy gamma-ray detectors with sources emitting photons in cascade
Author/Authors :
M. Korun، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
5
From page :
207
To page :
211
Abstract :
A method for measurement of peak and total efficiencies for point sources measured on low-energy gamma-ray detectors is presented. The efficiencies are measured with radionuclides emitting gamma-rays in a two-step cascade or in a crossover transition. One of the photons emitted in the cascade must have an energy which is much lower than the other in order to attain stronger attenuation as compared to the higher energy photon. The method is illustrated by measurements of the efficiencies with 179Ta and 57Co.
Keywords :
Gamma-ray spectrometry , attenuation , Efficiency calibration , Coincidence summing , Close geometry
Journal title :
Applied Radiation and Isotopes
Serial Year :
2004
Journal title :
Applied Radiation and Isotopes
Record number :
541586
Link To Document :
بازگشت