Title of article
Temperature dependence of spurious pulses in use of plastic scintillation detectors
Author/Authors
Yasushi Kawada، نويسنده , , Jun Ito، نويسنده , , Qiu-Wei Wang، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
5
From page
403
To page
407
Abstract
Pulse-height and time spectra of afterpulses from a scintillation detector with NE102A scintillator sandwiching a 60Co source were observed at several different temperatures. From time analysis with a slow amplifier followed by a conventional TAC, spurious pulses are grouped into two types; time-dependent afterpulses and random noise. The intensity of afterpulses decreased considerably with increasing temperature, while the intensity of random noise pulses increased abruptly above 50°C. Experiments, in which the temperatures of scintillator sheets and of the photomultiplier tube were changed separately, suggested that the time-dependent afterpulses were produced in the scintillator itself, while the random component arose from photomultiplier tube.
Keywords
Spurious pulse , Afterpulse , Photomultiplier tube , temperature dependence , plastic scintillator
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541617
Link To Document