Title of article
Ultra low-level measurements of actinides by sector field ICP-MS
Author/Authors
F. Pointurier، نويسنده , , N. Baglan، نويسنده , , P. Hémet، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
561
To page
566
Abstract
In the present work, a double-focusing sector field inductively coupled plasma-mass spectrometer was optimised for ultra trace and isotopic analyses of actinide long-lived isotopes in low concentration solutions of the fg ml−1 to the ng ml−1 range. Sensitivities of about 3 GHz/(μg ml−1), with as low a background as 0.1 cps, were obtained for U using a conventional concentric pneumatic nebuliser. Detection limits are below the fg range for 239Pu and 240Pu. With natural U, a precision lower than 0.5% RSD is currently obtained for 235U/238U isotopic ratio at the 200 pg ml−1 level.
Keywords
U , pu , Long-lived-isotopes , Trace analyses , Environmental samples , Double focusing sector field ICP-MS
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541645
Link To Document