• Title of article

    Efficiency calibration of high volume samples using the GESPECOR software

  • Author/Authors

    O. Sima، نويسنده , , I. L. Cazan، نويسنده , , L. Dinescu، نويسنده , , Eugene D. Arnold، Jr نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    5
  • From page
    123
  • To page
    127
  • Abstract
    In this paper the Monte Carlo simulation program GESPECOR is applied for the computation of the efficiency of an HPGe gamma-spectrometry system used for the measurement of high volume samples. The detector characterization required for Monte Carlo simulation is achieved by a trial and error procedure using the efficiency values measured with point sources placed in several positions. After detector characterization, the efficiency values appropriate for the assessment of 220 l waste drums are computed and compared with experimental values.
  • Keywords
    Gamma-ray spectrometry , Ge detector , Detector characterization , Efficiency calibration , Radioactive waste assessment
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number