Title of article
Efficiency calibration of high volume samples using the GESPECOR software
Author/Authors
O. Sima، نويسنده , , I. L. Cazan، نويسنده , , L. Dinescu، نويسنده , , Eugene D. Arnold، Jr نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
5
From page
123
To page
127
Abstract
In this paper the Monte Carlo simulation program GESPECOR is applied for the computation of the efficiency of an HPGe gamma-spectrometry system used for the measurement of high volume samples. The detector characterization required for Monte Carlo simulation is achieved by a trial and error procedure using the efficiency values measured with point sources placed in several positions. After detector characterization, the efficiency values appropriate for the assessment of 220 l waste drums are computed and compared with experimental values.
Keywords
Gamma-ray spectrometry , Ge detector , Detector characterization , Efficiency calibration , Radioactive waste assessment