Title of article :
Development of radon calibration chamber at KRISS
Author/Authors :
Jong-Man Lee، نويسنده , , Kil Hoon Ahn، نويسنده , , Ha Seok Chai، نويسنده , , Tae Soon Park، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
5
From page :
237
To page :
241
Abstract :
Radon calibration chamber has been constructed to test and calibrate radon and radon progeny detectors at various environmental conditions and to study the characteristic behavior of radon decay products. The size of main room in the chamber is 3.2 m×3.3 m×2.4 m (25.3 m3) and the total volume of the chamber is 39.1 m3. The maximum concentration of radon in the chamber can be maintained up to 250 kBq m−3 and temperature and humidity are controlled in the ranges of (10–40)°C and (30–80)%, respectively. The equilibrium factor and unattached fraction of radon progeny are determined at the radon concentration of 2 kBq m−3.
Keywords :
RADON , Radon progeny , Radon chamber , Equilibrium factor
Journal title :
Applied Radiation and Isotopes
Serial Year :
2004
Journal title :
Applied Radiation and Isotopes
Record number :
541740
Link To Document :
بازگشت