Title of article :
Performance comparison of neutron and X-ray sensitive photo-stimulated imaging plates
Author/Authors :
Hisao Kobayashi، نويسنده , , Masanori Satoh، نويسنده , , Masahiro Etoh، نويسنده , , Kohichi Ogura، نويسنده , , Hironobu Yanagie، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
6
From page :
573
To page :
578
Abstract :
Statistical characteristics of photo-stimulated luminescent (PSL) signals are studied using thermal neutrons, X- and γ-rays. It is shown that the statistical fluctuation of PSL is described simply by the PSL1 value itself and is independent of radiation type, except for short-range high-LET particles. However, it is expected that the spatial resolution should be affected if the range of electrons or secondaries were much longer than the instrumental resolution (pixel size). When output PSL signals are compared between X- and γ-rays, the PSL values were proportional to the exposed air dose, independent of X-ray energy. However, signals from 60Co γ-rays were 1% of 22–40 kV X-ray output at the same dose.
Keywords :
X-rays , Imaging plate , Air-dose , ?-rays , sensitivity , Linearity , Statistical character , Single radiation detection , neutrons
Journal title :
Applied Radiation and Isotopes
Serial Year :
2004
Journal title :
Applied Radiation and Isotopes
Record number :
541793
Link To Document :
بازگشت