Title of article :
Implantable self-powered detector for on-line determination of neutron flux in patients during NCT treatment
Author/Authors :
M. E. Miller، نويسنده , , L. E. Mariani، نويسنده , , M. L. Sztejnberg Gonçalves-Carralves، نويسنده , , M. Skumanic، نويسنده , , S. I. Thorp، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
5
From page :
1033
To page :
1037
Abstract :
A novel system to determine thermal neutron flux in real time during NCT treatments was developed in the National Atomic Energy Commission of Argentina. The system is based on a special self-powered detector that can be implanted in patients owing to its small size and biocompatibility. High voltage is not required to operate this kind of detectors, which is a considerable advantage in terms of medical uses. By choosing the appropriate materials, it was possible to obtain a prototype with thermal neutron sensitivity providing for an adequate signal level in typical NCT thermal fluxes. It was also possible to minimize gamma response in order to neglect its contribution.
Keywords :
SPND , NCT , On-line neutron flux measurement , Implantable detector
Journal title :
Applied Radiation and Isotopes
Serial Year :
2004
Journal title :
Applied Radiation and Isotopes
Record number :
541869
Link To Document :
بازگشت