Title of article :
EPR dosimetry using commercial glasses for high gamma doses
Author/Authors :
Maria Inês Teixeira، نويسنده , , Gilberto M. Ferraz، نويسنده , , Linda V.E Caldas، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
365
To page :
370
Abstract :
Commercial transparent and colored (bronze, brown, and green) glasses were studied as possible dosimeters for high gamma doses using electronic paramagnetic resonance (EPR). All EPR spectra showed the characteristic Fe3+ signals, g=4.27 and 2.01. The signal at g=2.01 presented a more useable behavior for the calibration curve. All samples showed their usefulness as high dose dosimeters.
Keywords :
GLASS , Dosimeter , EPR , High doses
Journal title :
Applied Radiation and Isotopes
Serial Year :
2005