Abstract :
High-frequency electromagnetic fields in the 120 kHz band emitted from card readers for access control systems caused abnormally high doses on electronic pocket dosimeters (EPDs). All EPDs recovered their normal performance by resetting after the exposure ceased. The electric and magnetic immunity levels of the EPDs were estimated by using the distances needed to prevent electromagnetic interference.
Keywords :
Security system , Electronic pocket dosimeter (EPD) , electromagnetic interference , Access control system , electromagnetic immunity , Card reader