Title of article :
Influence of dimensional stability of impression materials on the probability of acceptance of a prosthetic restoration
Author/Authors :
M. Pamenius، نويسنده , , N. G. Ohlson، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
The accuracy of rubber impression materials was evaluated by taking impressions of a steel model, pouring it with stone and comparing certain measurements between steel and stone models. It is assumed that acceptance is achieved when the difference in measurements is within permissible limits, defined by clinical criteria. A mathematical formulation of these criteria will lead to the concept of probability of acceptance. The probability is affected when the limits are changed. The particular case of satisfying two acceptance criteria that contain variables which are not statistically independent s considered, leading to the application of the bivariate normal distribution in the evaluation of acceptance probability.
Keywords :
accuracy , Dental materials , elastomeric impressions , Acceptance criteria , bivariate normalldistribution
Journal title :
Biomaterials
Journal title :
Biomaterials