Title of article
Release of foreign bodies (particles) by clinical use of intravenous infusion sets
Author/Authors
Henrik Madsen، نويسنده , , Ole Winding، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
4
From page
663
To page
666
Abstract
In clinical practice, stripping the plastic tubes of intravenous (i.v.) infusion sets with a scissor blade is a commonly used method for re-establishing flow in malfunctioning i.v. sets. The present investigation concludes that this procedure results in release of plastic particles from the luminal wall of the tube. Particles are subsequently flushed into the patient. The average amount of particles released exclusively from the i.v. infusion sets under these circumstances may exceed the standards for acceptable particle content per millilitre in large volume i.v. injectable fluids, according to the British Pharmacopoeia (1980) and the United States Pharmacopoeia (1990).
Keywords
Intravenous infusion sets , contamination , Particles , Stripping
Journal title
Biomaterials
Serial Year
1996
Journal title
Biomaterials
Record number
542442
Link To Document