• Title of article

    Release of foreign bodies (particles) by clinical use of intravenous infusion sets

  • Author/Authors

    Henrik Madsen، نويسنده , , Ole Winding، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    4
  • From page
    663
  • To page
    666
  • Abstract
    In clinical practice, stripping the plastic tubes of intravenous (i.v.) infusion sets with a scissor blade is a commonly used method for re-establishing flow in malfunctioning i.v. sets. The present investigation concludes that this procedure results in release of plastic particles from the luminal wall of the tube. Particles are subsequently flushed into the patient. The average amount of particles released exclusively from the i.v. infusion sets under these circumstances may exceed the standards for acceptable particle content per millilitre in large volume i.v. injectable fluids, according to the British Pharmacopoeia (1980) and the United States Pharmacopoeia (1990).
  • Keywords
    Intravenous infusion sets , contamination , Particles , Stripping
  • Journal title
    Biomaterials
  • Serial Year
    1996
  • Journal title
    Biomaterials
  • Record number

    542442