Title of article :
Characterization of electrolytic ZrO2 coating on Co–Cr–Mo implant alloys of hip prosthesis
Author/Authors :
S. K. Yen، نويسنده , , M. J. Guo، نويسنده , , H. Z. Zan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
9
From page :
125
To page :
133
Abstract :
An electrolytic Zr(OH)4 gel has been coated on ASTM F-75 Co–Cr–Mo alloy specimens in 0.0625 M ZrO(NO3)2 solution with pH=2.2 at a current density of 2 mA/cm2. After annealing at 623–973 K for 120 min in air, the ZrO2-coated specimen was evaluated by electrochemical polarization in Hankʹs solution, wear tests with UHMWPE (Ultra-high molecular-weight polyethylene) under a load stress of 50 MPa, scratch tests, surface morphology observations, and XRD analysis. The ZrO2-coated specimen annealed at 773 K for 120 min revealed a good adhesion of 610 MPa on Co–Cr–Mo substrate, a lower wear loss of UHMWPE and a higher protection potential than the uncoated specimen in Hankʹs solution. A monoclinic structure with preferred orientation parallel to the sheet plane was detected at 623 K T 673 K and a tetragonal structure of ZrO2 was detected at T 773 K. Then a monoclinic structure with random orientation and a tetragonal structure were mixed at T 973 K.
Keywords :
Electrolytic ZrO2coating , Co}Cr}Mo implant alloy , hip prosthesis
Journal title :
Biomaterials
Serial Year :
2001
Journal title :
Biomaterials
Record number :
543708
Link To Document :
بازگشت