Title of article :
New methods for the assessment of in vitro and in vivo stress cracking in biomedical polyurethanes
Author/Authors :
Darren J. Martin، نويسنده , , Laura A. Poole Warren، نويسنده , , Pathiraja A. Gunatillake، نويسنده , , Simon J. McCarthy، نويسنده , , Gordon F. Meijs، نويسنده , , Klaus Schindhelm، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
6
From page :
973
To page :
978
Abstract :
This article describes a new test method for the assessment of the severity of environmental stress cracking of biomedical polyurethanes in a manner that minimizes the degree of subjectivity involved. The effect of applied strain and acetone pre-treatment on degradation of Pellethane 2363 80A and Pellethane 2363 55D polyurethanes under in vitro and in vivo conditions is studied. The results are presented using a magnification-weighted image rating system that allows the semi-quantitative rating of degradation based on distribution and severity of surface damage. Devices for applying controlled strain to both flat sheet and tubing samples are described. The new rating system consistently discriminated between the effects of acetone pre-treatments, strain and exposure times in both in vitro and in vivo experiments. As expected, P80A underwent considerable stress cracking compared with P55D. P80A produced similar stress crack ratings in both in vivo and in vitro experiments, however P55D performed worse under in vitro conditions compared with in vivo. This result indicated that care must be taken when interpreting in vitro results in the absence of in vivo data.
Keywords :
Polyurethane , ESC , biostability , Pellethane
Journal title :
Biomaterials
Serial Year :
2001
Journal title :
Biomaterials
Record number :
543800
Link To Document :
بازگشت