Title of article
Mechanical properties of artificial tracheas composed of a mesh cylinder and a spiral stent
Author/Authors
Satoshi Kawaguchi، نويسنده , , Tatsuo Nakamura، نويسنده , , Yasuhiko Shimizu، نويسنده , , Toshiro Masuda، نويسنده , , Toshikazu Takigawa، نويسنده , , Yu Liu، نويسنده , , Hiroki Ueda، نويسنده , , Takashi Sekine، نويسنده , , Kazuya Matsumoto، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
6
From page
3085
To page
3090
Abstract
Much work has been done on the materials used for mesh-type artificial tracheas, but a precise mechanical evaluation of these structures has not yet been performed. In the present study, we determined the mechanical properties of typical mesh-type artificial tracheas and compared them with those of native trachea. Four types of artificial trachea were made and used for the mechanical tests. The basic frame of all the specimens was composed of a mesh cylinder and a spiral stent. The specimen whose mesh was sealed with collagen sponge showed almost the same behavior in the force-strain curve under compression, suggesting that collagen sealing has little effect on mechanical properties. Agreement between measured and estimated mechanical properties was good, especially in the low strain region, suggesting that artificial tracheas can be designed in terms of mechanical properties by mainly considering the basic frame structure.
Keywords
Artificial trachea , Canine trachea , Tensile test , Mechanical property , Young’s modulus , Compression test
Journal title
Biomaterials
Serial Year
2001
Journal title
Biomaterials
Record number
544039
Link To Document