• Title of article

    Mechanical and histomorphometric evaluations of titanium implants with different surface treatments inserted in sheep cortical bone

  • Author/Authors

    Gianluca Giavaresi، نويسنده , , Milena Fini، نويسنده , , Alberto Cigada، نويسنده , , Roberto Chiesa، نويسنده , , Gianni Rondelli، نويسنده , , Lia Rimondini، نويسنده , , Paola Torricelli، نويسنده , , Nicol? Nicoli Aldini، نويسنده , , Roberto Giardino، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    12
  • From page
    1583
  • To page
    1594
  • Abstract
    Improvement of the implant–bone interface is still an open problem and the interest in chemical modification of implant surfaces for cementless fixation has grown steadily over the past decade. Mechanical and histomorphometric investigations were performed at different times on implants inserted into sheep femoral cortical bone to compare the in vivo osseointegration of titanium screws ( 3.5×7 mm length) with different surface treatments. After 8 weeks of implantation, the push-out force of anodized and hydrothermally treated implants (ANODIC) was significantly higher than that of machined implants (MACH) (36%, p<0.0005), whereas a decrease of 39% was observed for acid-etched implants (HF) when compared to other surface treatments. After 12 weeks of implantation, the push-out force values of HF implants were still significantly lower than those observed for MACH (−19%, p<0.01) and hydroxyapatite vacuum plasma-sprayed implants (HAVPS, −25%, p<0.0005), and the highest push-out force was found in HAVPS (p<0.001) implants. After 8 and 12 weeks of implantation, the AI of HF implants was significantly (p<0.05) lower ( −25%) than that of MACH, HAVPS and ANODIC implants. In conclusion, results appear to confirm that there are no specific differences between ANODIC and HAVPS implants in terms of behavior. Moreover, although MACH implants show some surface contaminating agents, they appear to ensure good osseointegration within 12 weeks both mechanically and histomorphometrically, as do ANODIC and HAVPS implants. However, further studies are required to investigate bone hardness and mineralization around implants.
  • Keywords
    Titanium , sheep , Surface treatments , Histomorphometry , Mechanical testing
  • Journal title
    Biomaterials
  • Serial Year
    2003
  • Journal title
    Biomaterials
  • Record number