Title of article :
Loss of ions in cavity ionization chambers
Author/Authors :
N. Takata، نويسنده , , N.T. Tran، نويسنده , , E. Kim، نويسنده , , P. Marsoem، نويسنده , , T. Kurosawa، نويسنده , , Y. Koyama، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
4
From page :
805
To page :
808
Abstract :
Ion losses due to initial recombination, volume recombination, and back diffusion were each determined by measurements and calculations for different size cylindrical ionization chambers and spherical ionization chambers. By measuring signal currents from these ionization chambers irradiated with 60Co gamma rays, two groups of ion losses were obtained. (Group 1) Ion loss due to initial recombination and diffusion, which changes proportionally to the inverse of the voltage applied to the ionization chambers; (and group 2) ion loss due to volume recombination, which changes proportionally to the inverse of the square of the applied voltage. The diffusion loss was obtained separately by computing electric field distributions in the ionization chambers. It was found that diffusion loss is larger than initial recombination loss for the cylindrical ionization chambers and vise versa for the spherical ionization chambers.
Keywords :
ionization chamber , recombination , diffusion , Signal current , Volume , Initial , Ion loss
Journal title :
Applied Radiation and Isotopes
Serial Year :
2005
Journal title :
Applied Radiation and Isotopes
Record number :
547854
Link To Document :
بازگشت