Title of article :
Comparison of different methods for radiochemical purity testing of [99mTc-EDDA-HYNIC-D-Phe1,Tyr3]-Octreotide
Author/Authors :
Elisabeth von Guggenberg، نويسنده , , Barbara Penz، نويسنده , , Georg Kemmler، نويسنده , , Irene Virgolini، نويسنده , , Clemens Decristoforo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
7
From page :
194
To page :
200
Abstract :
[99mTc-EDDA-HYNIC-D-Phe1,Tyr3]-octreotide (99mTc-EDDA-HYNIC-TOC) is an alternative radioligand for somatostatin receptor (SSTR) scintigraphy of neuroendocrine tumours. In order to allow a rapid and accurate determination of the quality in the clinical routine the aim of this study was to evaluate different methods of radiochemical purity (RCP) testing. Three different methods of RCP testing were compared: high-performance liquid chromatograhpy (HPLC), thin layer chromatography (TLC) and minicolumn (Sep-Pak purification=SPE). HPLC was shown to be the most effective method for the quality control. The use of TLC and SPE is only recommended after sufficient practical labelling experience.
Keywords :
technetium-99m , N , HPLC , TLC , Tyr3]-Octreotide , Radiochemical purity
Journal title :
Applied Radiation and Isotopes
Serial Year :
2006
Journal title :
Applied Radiation and Isotopes
Record number :
547886
Link To Document :
بازگشت